HT701
EN - 23
6.1.1. Electrical characteristics
Conversion: TRMS
Sampling frequency: 3 times per second
Temperature coefficient: 0.15x(accuracy) /°C, <18°C or >28°C
NMRR Normal Mode Rejection Ratio: > 50dB for DC quantities and 50/60Hz
CMRR Common Mode Rejection Ratio: >100dB from DC up to 60Hz (DCV)
> 60dB from DC, up to 60Hz (ACV)
6.1.2. Considered standards
Safety: IEC/EN 61010-1
IEC/EN61557-1, IEC/EN61557-2
Insulation: double insulation
Pollution level: 2
Overvoltage category: CAT IV 600V, CAT III 1000V
Max operating altitude: 2000m (6561ft)
6.1.3. General characteristics
Mechanical characteristics
Size (with shell): 207 (L) x 95 (W) x 52 (H) mm (8 x 4 x 2 inch)
Weight (batteries included): 630g (22 ounces)
Power supply
Battery type: 4 x 1.5V alkaline type AA IEC LR6
Low battery indication: symbol “ ” with battery voltage < approx. 4.8V
Battery life: approx. 80 hours
600 insulation test with new batteries at room
temperature (1M @ 1kV, duty cycle of 5s on
and 25s off)
Auto Power Off after 20 minutes of idleness (disabled)
Memory
Characteristics: max. 100 locations for each function
Display
Characteristics: 5 LCD, 10000 points, decimal sign, analogue
bargraph and autobacklight
Over range indication: “OL” or “-OL”
6.2. ENVIRONMENT
6.2.1. Environmental conditions for use
Reference temperature: 23° ± 5°C (73 ± 41°F)
Operating temperature: 0° ÷ 50°C (32° ÷ 122°F)
Allowable relative humidity: <80%RH
Storage temperature: -20° ÷ 60°C (-4 ÷ 140°F) (batteries not fitted)
Storage humidity: <80%RH (batteries not fitted)
This instrument satisfies the requirements of Low Voltage Directive 2006/95/EC
(LVD) and of EMC Directive 2004/108/EC