FULLTEST 3
EN - 88
TRMS Leakage current using current clamp (ILEAK)
Range Display range Resolution
Accuracy
(w/o clamp error)
Overvoltage protection
1000 mA
0.0 99.9 mA
0.1 mA
± (3% rdg. + 3 dgt)
One measurement
terminal grounded
100 1000 mA
1 mA
100.0 A
0.00 9.99 A
0.01 A
10.0 100.0 A
0.1 A
1000 A
0.0 99.9 A
0.1 A
100 1000 A
1 A
Input voltage range: 0 1 V AC
Input resistance: 1 M
Frequency range: 40 Hz 100 kHz (characteristics according to
DIN EN 61010-1 Annex A, fig. A.1 and VDE 0411 Part 1)
Clamp type: AC, output voltage 1 V / measuring range, type
HT96U (measuring ranges 1 A, 100 A, 1000 A)
LIM value (1000 mA range): Adjustable 0.1 ... 99.9 mA, 100 ... 1000 mA
LIM value (100.0 A range): Adjustable 0.1 ... 100.0 A
LIM value (1000 A range): Adjustable 1 ... 1000 A
TRMS Leakage Current on test socket (ILEAK)
Display range Resolution Accuracy Overvoltage protection
0.25 49.99 mA
0.01 mA
± (3% rdg. + 3 dgt)* CAT II 300 V
0.05 0.99 A
0.01 A
1.0 10.0 A
0.1 A
* Clamp error not included
Load current influence ± 0.01 mA / A
Limit value Adjustable 0.25 19.99 mA, 20.0 49.9 mA, 0.05 0.99 A,
1.0 10.0 A
Frequency range 40 Hz 100 kHz (characteristics according to
DIN EN 61010-1 Annex A, fig. A.1 and VDE 0411 Part 1)
Measurement method Differential
Mains cord polarity exchange Internally by using the L POS key
UUT mains ON switch test Yes, the test done by measuring L current, limit value 25 mA
Over-range limit (I
PE) Yes, the measurement will be automatically interrupted 10 s
after exceeding 10 A
Over-range limit (I
L) Yes, the measurement will be automatically interrupted 10 s
after exceeding 18 A
Pre-Tests - Grounding of PE terminal on wall schuko socket
- Status of ON/OFF mains switch on UUT (measurement of
power consumption on schuko test socket, limit value 6 VA)