Copyright HT ITALIA 2013 Release EN 1.00 - 29/07/2013 ENGLISH User manual
HT4011 EN – 94.3.2. AC voltage measurement CAUTION The maximum input AC voltage is 600V. Do not measure voltages exceeding the limits given
HT4011 EN – 104.3.3. Frequency and Duty Cycle measurement CAUTION When measuring frequency with leads, the maximum input AC voltage is 6
HT4011 EN – 114.3.4. Resistance measurement CAUTION Before attempting any resistance measurement, remove power from the circuit under test a
HT4011 EN – 124.3.5. Capacitance measurement CAUTION Before carrying out capacitance measurements on circuits or capacitors, cut off power s
HT4011 EN – 134.3.6. Continuity test and diode test CAUTION Before attempting any resistance measurement, remove power from the circuit under
HT4011 EN – 144.3.7. Temperature measurement in °C and °F CAUTION Do not put the temperature probe into contact with live surfaces. Voltage
HT4011 EN – 154.3.8. AC current measurement CAUTION Before attempting any measurement disconnect all the test leads from the circuit under t
HT4011 EN – 165. MAINTENANCE 5.1. GENERAL INFORMATION 1. The instrument you purchased is a precision instrument. While using and storing the i
HT4011 EN – 176. TECHNICAL SPECIFICATIONS 6.1. TECHNICAL CHARACTERISTICS Accuracy is calculated as ± [% reading + (number of digits) x resoluti
HT4011 EN – 18Frequency with leads (Autorange) Range Resolution Accuracy Sensitivity Overvoltage protection10.00Hz 49.99Hz 0.01Hz (1.5%rdg+2
HT4011 EN – 1Table of contents: 1. PRECAUTIONS AND SAFETY MEASURES ... 21.1. Pr
HT4011 EN – 196.3. ACCESSORIES PROVIDED Pair of 2mm test leads Adapter + K-type wire probe Carrying bag Batteries (not fitted) Use
HT4011 EN – 207. SERVICE 7.1. WARRANTY CONDITIONS This instrument is warranted against any material or manufacturing defect, in compliance with
HT4011 EN – 21. PRECAUTIONS AND SAFETY MEASURES The instrument has been designed in compliance with directive IEC/EN61010-1 relevant to electro
HT4011 EN – 31.2. DURING USE Please carefully read the following recommendations and instructions: CAUTION Failure to comply with the Cautio
HT4011 EN – 42. GENERAL DESCRIPTION The instrument carries out the following measurements: DC and AC voltage up to 600V AC current up to
HT4011 EN – 53. PREPARATION FOR USE 3.1. INITIAL CHECKS Before shipping, the instrument has been checked from an electric as well as mechanica
HT4011 EN – 64. OPERATING INSTRUCTIONS 4.1. INSTRUMENT DESCRIPTION 4.1.1. Description of the controls CAPTION: 1. Inductive clamp jaw 2. A
HT4011 EN – 74.2. DESCRIPTION OF FUNCTION KEYS 4.2.1. HOLD key Short pressing the “HOLD” key activates the function Data HOLD, i.e. the value
HT4011 EN – 84.3. DESCRIPTION OF ROTARY SWITCH FUNCTIONS 4.3.1. DC voltage measurement CAUTION The maximum input DC voltage is 600Vrms. Do