MACROTEST 5035
EN - 14
In case the mode R+TIMER or R-
TIMER was selected and a R+ or R-
higher than 99.9Ω was detected, the
instrument emits an intermittent
sound signal during the test, a long
sound signal at the end of the test
and displays the screen alongside.
CAL LOWΩ
o> 99.9 Ω
- - - mA 4 s
R+
TIMER
In case that:
R
MEASURED
-R
CALIBRATION
< 0Ω the
instrument displays the screen
alongside.
CAL LOWΩ
o 0.00 Ω
219 mA
UTO
SAVING: The tests can be stored pressing the SAVE key twice
(according to paragraph 5.1).
If the terminal voltage is higher than
10V, the instrument does not carry out
the test and displays the screen
alongside for 5 seconds after which, the
instrument displays the screen relative
to the precedent selected test mode
LOWΩ under AUTO mode.
CAL LOWΩ
o 13V
VOL AG
THIS RESULT CANNOT BE SAVED.
99.9Ω is the
maximum value.
ATTENTION:
value of R+ or R-
is too high.
Blinking CAL.
ATTENTION: R
MEASURED
-R
CALIBRATION
< 0
AUTO mode.
SA
E
ATTENTION: the
test was not
effected because
of voltage at the
terminal ends.
SAVE