HT9022
EN - 43
5.8. POSITION “INRUSH A ”: INRUSH CURRENT MEASUREMENT
WARNING
The maximum measurable DC or AC+DC current is 1000A. Do not
measure currents exceeding the limits given in this manual. Exceeding
these limits could result in electrical shocks to the user and damage to the
instrument.
We recommend holding the clamp respecting the safety area created by
the hand protection (see Fig. 3).
Currents <3A are zeroed.
By positioning the selector switch to “InRush A ”, the screen to
the side will appear.
The display shows the current settings for recording inrush
currents (see par. 5.2.4).
Mod Zero Run Mem
50Hz 01/10
10 A
Fix 15A 1/1
19/01-17:00:00
Fig. 14: inrush current measurement
Press key F1 (Mod) to open the drop-down menu shown on the
screen to the side. At each subsequent pressure of key F1, the
cursor will scroll through the available items, as follows:
50Hz: 50Hz inrush current measurement;
60Hz: 60Hz inrush current measurement;
400Hz: 400Hz inrush current measurement;
DC: DC inrush current measurement;
Mod Zero Run OK
50Hz 01/10
10 A
Fix 15A 1/1
19/01-17:00:00
Help: it displays the connection between instrument and
system;
Esc: it closes the drop-down menu.
Select the desired mode and press key F4 (OK) to confirm.
50Hz
60Hz
400Hz
DC
Help