7.1.1. Safety standards
Compliant with Standards: IEC / EN61010-1, IEC / EN61010-2 – 032
Technical documentation: IEC/EN61187
Safety of measuring accessories: IEC/EN61010-31
Insulation: Class 2, double insulation
Pollution level: 2
Max height: 2000m, indoor use
Overvoltage category: CAT IV 600V / CAT III 1000V to earth, max 1000V between
inputs
7.1.2. General characteristics
Characteristics of radio module
Radio: Bluetooth V2.0
Frequency: 2.4 GHz (2400-2483.5MHz)
Power: Class 2
Data rate: 57600 baud
Memory
Internal memory: 2Mbytes
Recordings
N°. max Log + Snapshot stores: 99
N°. max InRush stores: 20 (each with max 10 events)
Mechanical characteristics
Size: 252 (L) x 88 (W) x 44 (H) mm; 9.92 x 3.46 x 1.73 in
Weight (batteries included): approx 420g; 14.8 ounces
Jaw opening / Max cable size: 45mm; 1.77 in
Power supply
Battery type: 2 batteries x 1.5V LR 03 AAA
Battery life: approx. 53 hours of use in “W ” position
Auto power OFF: 5 min. with enabled function The display shows the following
symbol “
Display
Characteristics: graphic display 128x128 pixels
Sampling rate: 128 samples per period (base sampling)
Updating frequency: 1/s
7.2. ENVIRONMENT
7.2.1. Environmental conditions for use
Reference calibration temperature: 23° ± 5 °C
Operating temperature: 0 ÷ 40 °C
Allowable relative humidity: < 80%
Storage temperature: -10 ÷ 60 °C
Storage humidity: < 70%
This instrument satisfies the requirements of Low Voltage Directive 2006/95/EEC
(LVD) and of EMC Directive 2004/108/EEC
7.3. ACCESSORIES PROVIDED
Instrument
Pair of test leads
Pair of alligator clips
ISO9000 calibration certificate
User manual
Bag
Batteries