PVCHECK
EN - 17
5.4. M – SETTINGS FOR INSULATION MEASUREMENT
5.4.1. Instrument settings
1. Position the cursor onto M by using the arrow keys
(,) and confirm with ENTER. The display shows the
following screen:
15/05/12 15:34:26
Ins.Test : 1000 V
Ri min : 1.0 M
Mode : Field
Vtest - - - V - - - V
Ri(+) - - -
M
Ri(-) - - -
M
Rp - - -
M
Selection
M
2. Press the ENTER key. The instrument shows the option:
Settings:
3. Confirm with ENTER. The instrument shows the following
screen:
15/05/12 15:34:26
Ins. Test : 1000 V
Ri min : 1.0 M
Mode : Field
Vtest - - - V - - - V
Ri(+) - - -
M
Ri(-) - - -
M
Rp - - -
M
Settings
Selection
M
4. Use the arrow keys (,) for option setting, and select
the desired option by means of the arrow keys ( , ).
The following parameters can be set:
Ins. Test test voltage: 250, 500, 1000VDC
Mode operating modes: Field, Timer, String
Ri min minimum limit value for insulation resistance
Test time maximum value of test time (only for
TIMER mode) with a maximum value of 300s in steps
of 1s.
5. Press the SAVE key to save the settings made; the
message “Data saved” will be displayed for a few
seconds. Press the ESC/MENU key to exit without saving
and go back to the previous screen.
15/05/12 15:34:26
Ins. Test : 1000 V
Ri min : 1.00 M
Mode : TIMER
Test time: 10s
SAVE to store
M
CAUTION
The settings saved for test voltage also affect the settings for the insulation
measurement contained in the IVCK measurement (MENU IVCK)